CPS100 SERIES OPERATION MANUAL
SECTION 6:
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CPS100 Series Power Source Operation Manual
Page 126 of 231
For the test levels, there are three standard
product classes for the IEC 61000-4-13. They
are Class 1, Class 2 and Class 3. (Standard
modes) Test levels and durations for these
product classes are defined by the IEC 61000-
4-13 standard. There is also a User class,
which allows parameters to be changed.
(User mode).
To select the appropriate product class for
the unit under test, move the cursor to the
“Level =” field and press [Enter]. A drop
down list with product class selections is
displayed.
The following other test parameters are available:
•
Unom:
Nominal test voltage. Fixed at 230V in standard modes.
•
Frequency:
50Hz or 60Hz in standard modes.
For User class, set the Unom parameter as
desired. For classes 1 through 3, select the
required frequency – typically 50Hz.
Once set, press the
Save
soft key to store the
test setting in non-volatile memory
Note:
If the IEC 4-13 test setting is not
saved, it will be lost when the power
source is turned off.