The IEC 4-13 test option is suitable for pre-
compliance testing of products that require
CE marking.
To set up and run an IEC 4-13 test, select the
IEC 4-13
soft key from the IEC Standard
menu.
The IEC 4-13 Menu contains the following
available soft keys:
•
Flat Curve
•
Over Swing
•
Individual Harmonic
•
Inter-harmonics
•
Sweep in Frequency
These are all part of the IEC 61000-4-13 test
requirements.
IEC 4-13 Flat Curve Test
Press the
Flat Curve
soft key to display the
Flat Curve test screen. Available soft keys in
this screen are
Trigger
and
Edit
. Press the
Edit
soft key to set the required test
selections for this test.
There are three standard product classes for
the IEC 61000-4-13. They are Class 1, Class 2
and Class 3. (Standard modes) Test levels and
durations for these product classes are
defined by the IEC 61000-4-13 standard.
There is also a User class, which allows
parameters to be changed. (User mode).
To select the appropriate product class for the unit under test, move the cursor to the “Test
Level =” field and press [Enter]. A drop down list with product class selections is displayed.
The following other test parameters are available:
•
AMP:
Voltage amplitude in percent of nominal voltage. Fixed at 100% in
standard modes.
•
Unom:
Nominal test voltage. Fixed at 230V in standard modes.
•
Frequency:
50Hz or 60Hz in standard modes.