CPS100 SERIES OPERATION MANUAL
SECTION 7:
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CPS100 Series Power Source Operation Manual
Page 160 of 231
7.8.1
CCMODE Sub-system
The CCMODE sub-system controls constant current mode operation.
Command
Description
CCMODE:_<nr2>
Enables (1) or Disables (0) the Constant Current mode. Returns
OK of FALSE.
CCMODE?
Queries the Constant Current mode state. Returns 0 (Disabled)
or 1 (Enabled).
Table 7-9: CCMODE Sub-system Commands
7.8.2
LIST Sub-system
The LIST sub-system controls output transient sequence programming.
Command
Description
LIST:BASE:_<nr2>
Sets the base for the selected step as either Cycle (0) or Time
(1). Returns OK or FALSE.
LIST:COUNT:_<nr2>
Set the repeat time for the selected step. Range is 0 ~ 9999. A 0
value means an infinite loop. Returns OK or FALSE.
LIST:CYCLE?
Queries the repeat time of the current step.
LIST:DEGREE:_<nr2>
Sets the start angle for the selected step in degrees. Range is
0.0 ~ 359.9 degrees. Returns OK or FALSE.
LIST:FREQEND:_<nr2>
Sets the end frequency for the selected step in Hz. Range is
15.00 ~ 1000.00 for Standard models or 15.00 ~ 1200.00 for
Enhanced models. Returns OK or FALSE.
LIST:FREQSTART:_<nr2>
Sets the start frequency for the selected step in Hz. Range is
15.00 ~ 1000.00 for Standard models or 15.00 ~ 1200.00 for
Enhanced models. Returns OK or FALSE.
LIST:LOAD
Loads programmed list sequence to power source’s list
memory. Returns OK of FALSE.
LIST:SAVE
Save the list sequence to no-volatile memory. Returns OK or
FALSE.
Note: There is no way at present to recall the List file directly
after cycling power on the unit. Because the file is saved in the
Display LCD controller instead of DSP. To run a the list file after
power on, the data must be transferred from screen to DSP
first.
LIST:MODE:_<nr2>
Sets the running mode of the list to either Continuous (0) or
Step mode (1). Returns OK or FALSE.
LIST:NUMM:_<nr2>
Save the settings to a list. Range is 0 ~ 49.
LIST:RUN
Starts execution of a list sequence. Returns OK or FALSE.
SYST:MODE?
Queries the running status of the list file. Returns 1 is running
or 0 if completed.