CPS100 SERIES OPERATION MANUAL
SECTION 7:
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CPS100 Series Power Source Operation Manual
Page 161 of 231
Command
Description
LIST:STATUS?
Queries the status of the list sequence.
0 = Running, 1 = Completed.
The command LIST:STATUS? and SEQ:STATUS? Commands can
be used when the Trigger mode is set to STEP, that means
when LIST:MODE or LIST:STEPMODE are set to 1 (step).
Note: LIST:STATUS? and SEQ:STATUS? Commands query the
status of the current step only, not the status of the whole list
file.
Please using "SYS:MODE?" query the running status of the list
file, return 1 (running) or 0 (finished
LIST:STEP:ALL:_<nr2>,..,<nr2>
Sets all parameters for a list step at once. There are 18
parameters for each list step. The 18 parameters must all be
specified in the exact order shown below.
Parameter1=SEQ:LISTID
Parameter2=LIST:MODE
Parameter3=SEQ:STEPCOUNT
Parameter4=LIST:STEPNUM
Parameter5=LIST:STEPID
Parameter6=LIST:NUM -------
(
0~49
)
Parameter7=LIST:STEPMODE
Parameter8=LIST:COUNT
Parameter9=LIST:WAVEFORM
Parameter10=LIST:BASE
Parameter11=LIST:DEGREE
Parameter12=LIST:FREQSTART
Parameter13=LIST:FREQEND
Parameter14=LIST:VACSTART
Parameter15=LIST:VACEND
Parameter16=LIST:VDCSTART
Parameter17=LIST:VDCEND
Parameter18=LIST: TIME
Programmed list Example:
SEQ:STEPNUM: 1
SEQ:CYCLE: 1
LIST:STEP:ALL: 1,0,3,4,1,2,0,1,A,1,0.0,,50.00,50.00,220.0,220.0,0.0,0.0,1000.0
LIST:STEP:ALL: 1,0,3,4,2,2,0,1,A,1,0.0,50.00,50.00,0.0,0.0,0.0,0.0,1000.0
LIST:STEP:ALL: 1,0,3,4,3,2,0,1,A,1,0.0,50.00,50.00,220.0,220.0,0.0,0.0,2000.0
LIST:STEP:ALL: 1,0,3,4,4,2,0,1,A,1,0.0,50.00,50.00,0.0,0.0,0.0,0.0,1000.0
LIST:LOAD
<DELAY TIME>1000 /* allow time to load list data to controller
LIST:SAVE
LIST:RUN
LIST:STEP?
Queries the number of steps in the list. Range is 0 ~ 49.
LIST:STEPID:_<nr2>
Sets the number for the current list step. Range is 1 ~ 9.
Returns OK or FALSE.
LIST:STEPMODE:_<nr2>
Sets the run mode for the current list step as Continuous (0) or
Step mode (1). Returns OK or FALSE.