CPS100 SERIES OPERATION MANUAL
SECTION 6:
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CPS100 Series Power Source Operation Manual
Page 123 of 231
IEC 4-13 Over Swing Test
From the top level IEC 4-13 menu, press the
Over Swing
soft key to display the Over
Swing test screen.
Available soft keys in this screen are
Trigger
and
Edit
. Press the
Edit
soft key to set the
required test selections for this test.
There are three standard product classes for
the IEC 61000-4-13. They are Class 1, Class 2
and Class 3. (Standard modes) Test levels and
durations for these product classes are
defined by the IEC 61000-4-13 standard.
There is also a User class, which allows
parameters to be changed. (User mode).
To select the appropriate product class for
the unit under test, move the cursor to the
“Test Level =” field and press [Enter]. A drop
down list with product class selections is
displayed.
The following other test parameters are available:
•
Unom:
Nominal test voltage. Fixed at 230V in standard modes.
•
Frequency:
50Hz or 60Hz in standard modes.
•
Test Time:
Duration of the test. Range is from 0 ~ 9999. Zero means infinite
loop.
•
Harmonic 3:
Amplitude in percent