11
Calibration
Overview
All Adaptive Power Systems’ instruments are factory calibrated prior to shipment. The
recommended calibration interval for CPS100 Series instruments is one year (every 12
months).
Calibration equipment required.
The following equipment is required to perform calibration on any of the CPS100 Series
power sources.
Item Type
Range
Suggested Make & Model
1
DMM
300 Vac RMS min.
Keysight, Model 34465A Digital Multimeter,
6½ Digit
2
Precision Current
Shunt
50 Arms
1
, 0.01
Ohm
Precision Shunts, Model CS-50, 0.01W,
0.01% (or equivalent
http://www.ohm-labs.com/precision-
3
Adjustable
Resistive or
Programmable
Load
240V min.
Adaptive Power Systems, Model 3C, AC & DC
programmable load or equivalent..
https://adaptivepower.com/products/ac-dc-
Table 11-1: Required Calibration Equipment List
Note 1:
For lower power models, a suitable smaller precision shunt may be used.