CPS100 SERIES OPERATION MANUAL
SECTION 7:
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CPS100 Series Power Source Operation Manual
Page 153 of 231
Measurement Commands
Measurements commands are all query which return a measurement result.
Note:
These measurement commands apply to an individual CPS100 series power source.
For Master/Slave configuration measurement commands, refer also to Section
7.2.2, “Master / Slave Measurement Commands” on page 171.
Command
Description
MEAS:VOLT?
Returns the Voltage RMS value of the output in volts.
MEAS:VDC?
Returns the DC Voltage value of the output in volts.
MEAS:VAC?
Returns the AC Voltage value of the output in volts.
MEAS:I?
Returns the Current RMS value of the output current in amps.
MEAS:IDC?
Returns the DC Current value of the output current in amps.
MEAS:IAC?
Returns the AC Current value of the output current in amps.
MEAS:FREQ?
Returns the output frequency measurement in Hz.
MEAS:VPK?
Returns the peak voltage of the output measurement in volts.
MEAS:IPK?
Returns the peak current measurement at the output in amps.
MEAS:CF?
Returns the crest factor measurement of the output voltage
MEAS:IS?
Returns the inrush current of the output current in amps.
MEAS:POWER?
Returns the true power output measurement in watts.
MEAS:VAR?
Returns the reactive power output measurement in VAR.
MEAS:VA?
Returns the apparent power output measurement in volt-amps.
MEAS:PF?
Returns the Power Factor of the output power.
MEAS:ALL?
Returns all 14 measurements, Output State and Alarm code as a comma
separated data string.
The order in which measurement values are returned is as follows.
Index - Measurement
Index – Measurement
Index – Meas. / State
0 – Vrms
6 – Freq
12 – VAR
1 – Vac
7 – Vpk
13 – VA
2 – Vdc
8 – Ipk
14 – PF
3 – Irms
9 – CF
15 – OUTPUT STATE
4 – Iac
10 – Is
16 – ALARM CODE
5 – Idc
11 – PWR
Table 7-2: Measurement Commands