Command
Description
OUTPUT:IMPED:RES:_<nr2>
Sets the Zo_R resistance of the output impedance. (Zo_R)
in ohms. Range is 0.01 ~ 1.00. Returns OK or FALSE.
OUTPUT:IMPED:RES?
Queries the R setting of the output impedance (Zo_R) in
Ohms.
OUTPUT:IMPED:INDU:_<nr2>
Sets the Zo_L inductance of the output impedance. (Zo_L)
in mH. Range is 0.01 ~ 1.00. Returns OK or FALSE.
OUTPUT:IMPED:INDU?
Queries the L setting of the output impedance (Zo_L) in
mH.
OUTPUT:RANGE:_<nr2>
Sets the output voltage range as follows:
<nr2> 0
150V
1
300V
2
AUTO
Return signal is OK or FALSE.
OUTPUT:RANGE?
Queries the voltage range setting.
0 = 150V / 1 =300V / 2 = AUTO
OUTPUT:COUPLE:_<nr2>
Sets the output coupling.
<nr2> 0
AC
1
DC
2
AC+DC
Return signal is OK or FALSE.
OUTPUT:COUPLE?
Queries the output coupling setting.
0 = AC / 1 = DC / 2 = AC+DC
OUTPUT:IRANGE:_<nr2>
Sets the current measurement range.
<nr2> 0
High Range
1
Middle Range
2
Low Range
3
mA Range
4
Auto Range
OUTPUT:IRANGE?
Queries the current measurement range.
0 = High / 1 = Middle / 2 = Low / 3 = mA / 4 = Auto
OUTPUT:ISSTART:_<nr2>
Sets the delay time for testing surge current Is in msec.
Range is 0 ~ 999. Returns OK or FALSE.
OUTPUT:ISSTART?
Queries the delay time setting for the inrush current test.
OUTPUT:ISINTERVAL:_<nr2>
Sets the interval time for testing surge current Is in msec.
Range is 0 ~ 999. Returns OK or FALSE.
OUTPUT:ISINTERVAL?
Queries the interval time setting for the inrush current
test.