112
5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Refill the Foreline Pump
Standard foreline pump only
Materials needed
•
Foreline pump oil (6040-0621) – approximately 0.28 L required
•
Funnel (9301-6461)
•
Gloves, oil- and solvent-resistant
•
Screwdriver, flat-blade, large (8730-0002)
•
Drain plug O-ring (if required) (0905-1515)
•
Hex key to remove drain plug (5 mm for Pfeiffer pump, 8710-1838)
Procedure
1
Drain the foreline pump. (See
2
Reinstall the drain plug. If the old O-ring appears worn or damaged,
replace it.
3
Remove the propping object from under the pump motor.
4
Add foreline pump oil until the oil level in the window is near, but not
above, the upper line. The foreline pump requires approximately 0.28 L of
oil.
5
Wait a few minutes for the oil to settle. If the oil level drops, add oil to bring
the oil level near the upper line.
6
Reinstall the fill cap.
7
If necessary, slide the foreline pump back under the analyzer chamber.
The foreline pump may be located on the floor, on the lab bench next to or
behind the MSD, or under the analyzer chamber at the back of the MSD.
8
Pump down the MSD.
WA R N I N G
The foreline pump can cause burns if touched when operating. It has a safety shield
to prevent the user from touching it.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...