Analyzer Heaters and Radiators
The ion source and mass filter are housed in cylindrical aluminum tubes
called radiators (See
The radiators control the distribution of heat in the analyzer. They also
provide electrical shielding for analyzer components. The source heater and
temperature sensor are mounted in the source heater block. The mass filter
(quad) heater and temperature sensor are mounted on the mass filter
radiator. Analyzer temperatures can be set and monitored from the MSD Data
Acquisition software.
In selecting the temperatures to use, consider the following:
•
Higher temperatures help keep the analyzer clean longer.
•
Higher ion source temperatures result in more fragmentation and,
therefore, lower high-mass sensitivity.
After pumpdown, it takes at least 2 hours for the analyzer to reach thermal
equilibrium. Data acquired sooner may not be reproducible.
Recommended settings (for EI operation):
•
Ion source 230
C
•
Quadrupole 150
C
The GC/MSD interface, ion source, and mass filter (quad) heated zones
interact. The analyzer heaters may not be able to accurately control
temperatures if the setpoint for one zone is much lower than that of an
adjacent zone.
C A U T I O N
Do not exceed 200 °C on the quadrupole, or 350 °C on the ion source.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...