CI Troubleshooting
Reagent gas ion ratio is difficult to adjust or unstable
The interface tip seal is incorrectly placed, damaged, or missing.
Inspect the Isolation tip. If necessary, remove and reinstall it to ensure a good
seal with the CI ion source. Replace it if it is damaged. Install it if it is missing.
Residual air and water in the MSD or in the reagent gas supply lines
Run the methane pretune. Air will appear as a peak at
m/z
32 and excessive
water as a peak at
m/z
19 >
m/z
17. If either of these conditions is present,
purge the reagent gas supply lines and bake out the MSD. See
on page 149. Continued presence of a large peak at
m/z
32 may indicate an air leak. After correcting the problems, you may need
to clean the ion source.
Air leak
Run Methane Pretune (in PCI). Large peak at
m/z
32 indicates air in the
system. Check for and correct any leaks. See
The reagent gas supply is at the wrong pressure.
Check the regulator on the reagent gas supply. It should be adjusted to 20 psi
(140 kPa).
A leak in the reagent gas delivery path.
This is especially likely if you have set the methane flow much higher than
normal and the ratio is still too low.
Check the reagent gas path. Tighten fittings.
The CI ion source is dirty.
Clean the ion source.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...