26
5977B Series MSD Troubleshooting and Maintenance Manual
1
Introduction
General laboratory precautions
•
Avoid leaks in the carrier gas lines. Use leak-checking equipment to
periodically check for hydrogen leaks.
•
Eliminate from your laboratory as many ignition sources as possible (open
flames, devices that can spark, sources of static electricity, etc.).
•
Do not allow hydrogen from a high pressure cylinder to vent directly to
atmosphere (danger of self-ignition).
•
Use a hydrogen generator instead of bottled hydrogen.
Operating precautions
•
Turn off the hydrogen at its source every time you shut down the GC or
MSD.
•
Turn off the hydrogen at its source every time you vent the MSD (do not
heat the capillary column without carrier gas flow).
•
Turn off the hydrogen at its source every time shutoff valves in an MSD are
closed (do not heat the capillary column without carrier gas flow).
•
Turn off the hydrogen at its source if a power failure occurs.
•
If a power failure occurs while the GC/MSD system is unattended, even if
the system has restarted by itself:
1
Immediately turn off the hydrogen at its source.
2
Turn off the GC.
3
Turn off the MSD, and allow it to cool for 1 hour.
4
Eliminate
all
potential sources of ignition in the room.
5
Open the vacuum manifold of the MSD to atmosphere.
6
Wait at least 10 minutes to allow any hydrogen to dissipate.
7
Start up the GC and MSD as normal.
When using hydrogen
check the system for leaks to prevent possible fire and
explosion hazards based on local Environmental Health and Safety (EHS)
requirements. Always check for leaks after changing a tank or servicing the
gas lines. Always make sure the vent line is vented into a fume hood.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...