5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Replace the DP Fluid
Materials needed
•
Aluminum foil, clean
•
Cloths, clean, lint-free (05980-60051)
•
Container for old DP fluid
•
DP fluid, 18.5 mL (6040-0809) – Two bottles are required
•
Gloves
•
Oil- and solvent-resistant
•
Thermally insulated
Procedure
1
Remove the DP from the MSD. (See
Remove the O-ring assembly from the top of the DP.
2
Cover the top of the DP with aluminum foil (shiny side up).
3
Heat the DP at 60 °C for 15 minutes. (For 7820A or 7890 GC’s you can use
the GC oven.)
4
Pour the old DP fluid out the top of the pump. Even after heating, the
pump fluid pours very slowly.
5
Check the color of the pump fluid.
WA R N I N G
The pump and pump fluid will be hot. Wear protective gloves when you remove the
pump from the oven.
WA R N I N G
Treat the old pump fluid as hazardous. It may contain traces of toxic chemicals.
WA R N I N G
Methylene chloride is a hazardous solvent. Work in a fume hood and take all
appropriate precautions.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...