5977B Series MSD Troubleshooting and Maintenance Manual
2
General Troubleshooting
Temperature Symptoms
The MSD has three heated zones:
•
Ion source
•
Mass filter
•
GC/MSD interface
Each heated zone has a heater and temperature sensor. The ion source and
mass filter are powered and controlled by the MSD. The GC/MSD interface is
powered and controlled by the GC. For the 7820A Series GC’s, the heater is
either connected to the rear inlet thermal zone for the single inlet models or
connected to the manual valve thermal zone for dual inlet models.
Ion source will not heat up
•
High vacuum pump is off or has not reached normal operating conditions
†
•
Incorrect temperature setpoint
•
Ion source has not had enough time to reach temperature setpoint
•
Ion source heater cartridge is not connected
†
•
Ion source temperature sensor is not connected
†
•
Ion source heater failed (burned out or shorted to ground)
†
•
Ion source temperature sensor failed
†
•
Source power cable is not connected to the side board
†
•
MSD electronics are not working correctly
† These will cause an error message.
Mass filter (quad) heater will not heat up
•
High vacuum pump is off or has not reached normal operating conditions
†
•
Incorrect temperature setpoint
•
Mass filter has not had enough time to reach temperature setpoint
•
Mass filter heater cartridge is not connected
†
•
Mass filter temperature sensor is not connected
†
•
Mass filter heater failed (burned out or shorted to ground)
†
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...