Introduction
1
5977B Series MSD Troubleshooting and Maintenance Manual
19
Important Safety Warnings
There are several important safety notices to keep in mind when using the
MSD.
Many internal parts of the MSD carry dangerous voltages
If the MSD is connected to a power source, even if the power switch is off,
potentially dangerous voltages exist on:
•
The wiring between the MSD power cord and the AC power supply, the AC
power supply itself, and the wiring from the AC power supply to the power
switch.
With the power switch on, potentially dangerous voltages also exist on:
•
All electronics boards in the instrument
•
The internal wires and cables connected to these boards
•
The wires for any heater (oven, detector, inlet, or valve box)
Electrostatic discharge is a threat to MSD electronics
The printed circuit boards in the MSD can be damaged by electrostatic
discharge. Do not touch any of the boards unless it is absolutely necessary. If
you must handle them, wear a grounded wrist strap and take other antistatic
precautions. Wear a grounded wrist strap any time you remove the MSD right
side cover.
WA R N I N G
All these parts are shielded by covers. With the covers in place, it should be difficult
to accidentally make contact with dangerous voltages. Unless specifically
instructed to, never remove a cover unless the detector, inlet, or oven are turned off.
WA R N I N G
If the power cord insulation is frayed or worn, the cord must be replaced. Contact
your Agilent service representative.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...