General Maintenance
4
5977B Series MSD Troubleshooting and Maintenance Manual
99
To Separate the MSD from the 9000 GC
Materials needed
•
Screwdriver, T-20 Torx (8710-1615)
Procedure
Separate the MSD and 9000 GC
1
Remove the 9000 GC/MSD Tail. Refer to the
Agilent 5977B Series MSD
Operating Manual.
2
Power off the GC.
3
Using a T-20 Torx screwdriver, loosen the lock plate by turning the lock
plate screw clockwise.
WA R N I N G
Ensure the GC/MSD interface and the analyzer zones are cool (below 100 °C) before
you vent the MSD. A temperature of 100 °C is hot enough to burn skin; always wear
cloth gloves when handling analyzer parts.
WA R N I N G
If you are using hydrogen as a carrier gas, the carrier gas flow must be closed
before turning off the MSD power. If the foreline pump is off, hydrogen will
accumulate in the MSD and an explosion may occur. Before operating the MSD with
hydrogen carrier gas read the hydrogen safety information. (See
C A U T I O N
Ensure the GC heated zones and the GC/MSD interface are cool before turning off
carrier gas flow.
WA R N I N G
Ensure the GC/MSD interface and GC heated zones have cooled before you remove
the 9000 GC/MSD Tail.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...