General Troubleshooting
2
5977B Series MSD Troubleshooting and Maintenance Manual
61
Contamination
Contamination is usually identified by excessive background in the mass
spectra. It can come from the GC or from the MSD. The source of the
contamination can sometimes be determined by identifying the contaminants.
Some contaminants are much more likely to originate in the GC. Others are
more likely to originate in the MSD.
Contamination originating in the GC typically comes from one of these
sources:
•
Column or septum bleed
•
Dirty GC inlet
•
GC inlet liner
•
Contaminated syringe
•
Poor quality carrier gas
•
Dirty carrier gas tubing
•
Fingerprints (improper handling of clean parts)
Contamination originating in the MSD typically comes from one of the
following sources:
•
Air leak
•
Cleaning solvents and materials
•
Foreline pump oil (standard foreline pump)
•
Fingerprints (improper handling of clean parts)
on page 62 lists some of the more common contaminants, the ions
characteristic of those contaminants, and the likely sources of those
contaminants.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...