CI Troubleshooting
Air Leaks
How do I know if I have an air leak?
Large
air leaks can be detected by vacuum symptoms: loud gurgling noise
from the foreline pump, inability of the turbo pump to reach 95% speed, or, in
the case of smaller leaks, high pressure readings on the high vacuum gauge
controller.
The mass flow controller is calibrated for methane, and the high vacuum gauge
controller is calibrated for nitrogen, so measurements are not accurate in
absolute terms.
Familiarize yourself with the measurements on
your
system under operating
conditions. Watch for
changes
that may indicate a vacuum or gas flow
problem.
Always
look for small air leaks when setting up methane flow. Run the
methane
pretune, starting with a good PCI tune file (
on page 67). The
abundance of
m/z
19 (protonated water) should be less than 50% of
m/z
17 for
acceptable PCI performance. For NCI, the abundance of
m/z
19 (protonated
water) should be less than 25% that of
m/z
17. If the MSD was just pumped
down, look for the abundance of
m/z
19 to be decreasing.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...