5977B Series MSD Troubleshooting and Maintenance Manual
5
CI Maintenance
To Refill the CI Calibrant Vial
Materials needed
•
PFDTD calibrant (8500-8510)
Procedure
1
Set the reagent gas flow to
Gas Off
.
2
Vent the MSD.
3
Remove the capillary column from the GC/MSD interface.
4
Pull the MSD away from the GC to expose the calibration vial and valve.
See
“To Separate the MSD from the 7890 GC”
5
Loosen the collar holding the calibration vial in place. Do not remove the
collar.
6
Remove the calibrant vial. See
7
Fill the vial no higher than the bottom of the internal tube with fresh
PFDTD calibrant (8500-8510).
8
Replace the vial and tighten the collar.
9
Reposition the MSD next to the GC. See
“To Reconnect the MSD to the 7890
10
Reinstall the capillary column.
11
Pump down the MSD.
12
Purge the calibration valve. See
“To Purge the Calibration Valves”
C A U T I O N
Do
not
rinse the vial with any solvents.
Never
expose the inside of the vial to
chlorinated solvents or isopropyl alcohol or water — this will result in severe loss of CI
sensitivity.
C A U T I O N
After removing the calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in severe contamination of the ion source and damage to the filament and
electron multiplier.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...