5977B Series MSD Troubleshooting and Maintenance Manual
6
Vacuum System
Analyzer Chamber
The analyzer chamber (
) is where the analyzer operates. The
manifold is extruded and machined from an aluminum alloy. Large openings
in the side, front, and rear of the analyzer chamber are closed by plates.
O-rings provide the seals between the plates and the manifold. Ports in the
manifold and the plates provide attachment points for the Micro-Ion vacuum
gauge, calibration valve, vent valve, GC/MSD interface, and high vacuum
pump.
Diffusion pump version
The diffusion pump attaches, with a KF50 seal, to a baffle adapter that is
clamped to the bottom of the manifold. A vapor baffle helps prevent migration
of pump fluid vapor into the manifold. Cooling fins on the bottom of the
manifold keep the baffle cool so the vapor will condense on it.
Turbo pump version
The turbo pump and the mounting bracket for the turbo controller are
clamped directly to the manifold.
Figure 29
Analyzer chamber
Vacuum gauge baffle
GC/MSD interface
To high vacuum pump
Calibration and vent valves
Observation window
Side plate O-ring
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...