![Agilent Technologies 5977B Series Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608037.webp)
General Troubleshooting
2
5977B Series MSD Troubleshooting and Maintenance Manual
37
Troubleshooting Tips and Tricks
Rule 1: Look for what has been changed.
Many problems are introduced accidentally by human actions. Every time any
system is disturbed, there is a chance of introducing a new problem.
•
If the MSD was just pumped down after maintenance, suspect air leaks or
incorrect assembly.
•
If carrier gas or helium gas purifier were just changed, suspect leaks or
contaminated or incorrect gas.
•
If the GC column was just replaced, suspect air leaks or contaminated or
bleeding column.
Rule 2: If complex is not working, go back to simple.
A complex task is not only more difficult to perform but also more difficult to
troubleshoot. If you are having trouble detecting your sample, verify that
autotune is successful.
Rule 3: Divide and conquer.
This technique is known as half-split troubleshooting. If you can isolate the
problem to only part of the system, it is much easier to locate.
•
To determine whether an air leak is in the GC or the MSD, you can vent the
MSD, remove the column, and install the blank interface ferrule. If the leak
goes away, it was in the GC.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...