General Maintenance
4
5977B Series MSD Troubleshooting and Maintenance Manual
109
To Check the Foreline Pump Fluid
Wet foreline pumps only
Materials needed
•
Foreline pump oil (6040-0621)
•
Funnel (9301-6461)
•
Hex key to remove drain plug (5 mm for Pfeiffer pump, 8710-1838)
•
Screwdriver, flat-blade, to remove top fill cap
Procedure
Always replace the oil if it is dark or cloudy or due for replacement instead of
adding oil. (See
1
Examine the oil level window (
2
Note the two lines on the pump left of the window. The oil level should be
between the lines. The foreline pump oil should be almost clear. If the oil level
is near or below the lower line, add foreline pump oil. (See
on page 112, starting at step four.)
WA R N I N G
The foreline pump can cause burns if touched when operating. It has a safety shield
to prevent the user from touching it.
WA R N I N G
Never add oil while the foreline pump is on.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...