General Maintenance
4
5977B Series MSD Troubleshooting and Maintenance Manual
111
To Drain the Foreline Pump
Wet foreline pump only
Materials needed
•
Book or other solid object approximately 5 cm thick
•
Container for catching old pump oil, 500 mL
•
Gloves, oil- and solvent-resistant
•
Screwdriver, flat-blade, large (8730-0002)
•
Hex key to remove drain plug (5 mm for Pfeiffer pump, 8710-1838)
Procedure
1
Vent the MSD.
2
If necessary, slide the foreline pump to a safe, accessible location.
The foreline pump may be located on the floor, on the lab bench next to or
behind the MSD, or under the analyzer chamber at the back of the MSD.
3
4
Place a container under the drain plug.
5
Remove the drain plug. Allow the pump oil to drain out. The oil drains
faster if it is still warm.
If necessary, you can place a book or other object under the pump motor to
tilt it up slightly.
6
Replace the drain plug after draining the oil.
WA R N I N G
The foreline pump can cause burns if touched when operating. It has a safety shield
to prevent the user from touching it.
WA R N I N G
The old pump oil may contain toxic chemicals. Treat it as hazardous waste.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...