![Agilent Technologies 5977B Series Скачать руководство пользователя страница 216](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608216.webp)
216
5977B Series MSD Troubleshooting and Maintenance Manual
8
Electronics
•
Generates ±280 VDC (nominal) power for main board lens amplifiers and
side board DC amplifiers.
•
Supplies and controls the power for the ion source and quadrupole (mass
filter) heaters.
•
Provides 24 VDC power for the cooling fans.
Signal amplifier board
The signal amplifier board amplifies the output of the detector. It produces an
output voltage of 0 to 10 V DC, proportional to the logarithm of the input
current of 3 picoamps to 50 microamps.
An analog-to-digital converter converts the amplifier output voltage to digital
information. The LAN/MSD control card converts the data into abundance
counts proportional to the detector signal current.
Lens driver board for HES
The lens driver board provides the lens voltages for the extractor, post
extractor 1, and post extractor 2 lenses.
AC board
The AC board is mounted on the opposite side of the electronics panel from
the LAN/MSD control card. The AC board is also sometimes called the power
distribution/vacuum control board. It performs these functions:
•
Provides input voltage transparency for the MSD.
•
Distributes AC line power to the AC/DC power supply, the foreline pump,
and the turbo pump controller.
•
Turns the calibration valve on or off as directed by the main board.
•
Provides the voltage for the calibration valve.
•
Provides a logic interface to turbo controller.
•
Controls the diffusion pump:
•
Controls the foreline gauge.
•
Turns on the diffusion pump once the foreline pressure is low enough, as
directed by the main board.
•
Regulates the AC power to the diffusion pump heater.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...