Vacuum System
6
5977B Series MSD Troubleshooting and Maintenance Manual
159
The standard foreline pump is a two-stage rotary-vane pump. An optional dry
pump is also available. The pump turns on when the MSD power is turned on.
The foreline pump has a built-in antisuckback valve to help prevent
backstreaming in the event of a power failure.
The foreline pump can be placed under the analyzer chamber at the rear of the
MSD (with the exhaust outlet to the rear) or on the floor below the MSD.
An oil trap (not shown) is available for the standard pump that can be used to
filter pump oil out of the foreline pump exhaust. This trap stops
only
pump
oil. Do not use the trap if you are analyzing toxic chemicals or using toxic
solvents or if you have a CI MSD. Instead, install an 11-mm id hose to remove
the exhaust from your lab.
A window (sight glass) in the front of the standard foreline pump shows the
level of the foreline pump oil. There are two marks next to the window. The
level of the pump oil should never be above the upper mark or below the lower
mark. If the level of pump oil is near the lower mark, add foreline pump oil.
WA R N I N G
The oil trap supplied with the standard foreline pump stops only foreline pump oil.
It does not trap or filter out toxic chemicals. If you are using toxic solvents or
analyzing toxic chemicals, remove the oil trap. Do not use the trap if you have a
CI MSD. Install a hose to take the foreline pump exhaust outside or to a fume hood.
C A U T I O N
Do not place the foreline pump near any equipment that is sensitive to vibration.
C A U T I O N
The ballast control knob controls the amount of air allowed into the pump. Keep the
ballast control closed (fully clockwise) at all times, except when ballasting the pump.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...