General Troubleshooting
2
5977B Series MSD Troubleshooting and Maintenance Manual
45
Mass Spectral Symptoms
This section describes symptoms you might observe in mass spectra. Some of
these symptoms will appear in the mass spectra of samples. Others you will
observe only in a tune report. Some of these symptoms have causes that can be
corrected by the operator. Others, however, require service by an Agilent
Technologies service representative.
The following symptoms listed under Chromatic symptoms also apply to mass
spectra:
•
If sensativity is low
•
If repeatability is poor
No peaks
•
Ion source cables not connected
•
Bad connections to or from the detector
•
HED power supply output cable has failed
•
Other electronics failure
•
Incorrect tune file (inappropriate parameters)
Isotopes are missing or isotope ratios are incorrect
•
Peaks are too wide or too narrow
•
Scan speed is too high (scan mode)
•
Dwell time is too short (SIM mode)
•
Electron multiplier voltage is too high
•
Repeller voltage is too high
•
High background
•
Dirty ion source
High background
•
Pressure in the analyzer chamber is too high
•
Air leak
•
Contamination
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...