6.
Input
DC
Offset
Adjustment
The
purp ose
of
this
pro cedure
is
to
adjust
the
DC
Bias
oset
v
alue.
Note
HP
VEE
Adjustmen
t
Program
is
not
required
in
this
adjustmen
t.
F
ollo
w
the
pro cedures
b elow
and
adjust
prop er
item
using
the
in
ternal
adjustmen
t
test.
Required
Equipment
None
Procedure
1.
T
urn
the
4395A
OFF.
2.
T
o
gain
access
to
the
adjustmen
t
comp onen
ts,
remov
e
the
b ottom
co
v
er
of
the
4395A.
3.
T
urn
the
4395A
ON.
4.
Press
4
System
5,
NNNNNNNNNNNNNN
MORE
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
TESTS
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADJUSTMENT
TESTS
to
display
the
in
ternal
adjustmen
t
tests
men
u.
5.
Press
4
6
5,
4
0
5
or
use
4
*
5/ 4
+
5
to
select
the
the
in
ternal
adjustmen
t
test
60:
A9
A
ON
ADJ.
6.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXECUTE
TEST
to
run
the
test.
7.
Adjust
\ADJ-A-ON"
on
the
A9
Input
Multiplexer
Board
un
til
pass
is
display
ed
on
the
screen.
8.
Press
NNNNNNNNNNNNNN
CONT
to
exit
the
adjustmen
t.
9.
Press
4
6
5,
4
1
5
or
use
4
*
5
to
select
the
in
ternal
adjustmen
t
test
61:
A9
B
ON
ADJ.
10.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXECUTE
TEST
to
run
the
test.
11.
Adjust
\ADJ-B-ON"
on
the
A9
Input
Multiplexer
Board
un
til
pass
is
display
ed
on
the
screen.
12.
Press
NNNNNNNNNNNNNN
CONT
to
exit
the
adjustmen
t.
13.
Press
4
6
5,
4
2
5
or
use
4
*
5
to
select
the
in
ternal
adjustmen
t
test
62:
A9
R
OFF
ADJ.
14.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXECUTE
TEST
to
run
the
test.
15.
Adjust
\ADJ-R-OFF"
on
the
A9
Input
Multiplexer
Board
un
til
pass
is
display
ed
on
the
screen.
16.
Press
NNNNNNNNNNNNNN
CONT
to
exit
the
adjustmen
t.
17.
Press
4
6
5,
4
3
5
or
use
4
*
5
to
select
the
in
ternal
adjustmen
t
test
63:
A9
A
OFF
ADJ.
18.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXECUTE
TEST
to
run
the
test.
19.
Adjust
\ADJ-A-OFF"
on
the
A9
Input
Multiplexer
Board
un
til
pass
is
display
ed
on
the
screen.
20.
Press
NNNNNNNNNNNNNN
CONT
to
exit
the
adjustmen
t.
5-16
Adjustments
Содержание 4395A
Страница 10: ......
Страница 26: ......
Страница 34: ......
Страница 77: ...Figure 2 17 B R Magnitude Ratio Phase Dynamic Accuracy Test Setup 2 Performance Tests 2 43 ...
Страница 167: ...Figure 5 1 Adjustment Hardware Setup Adjustments 5 5 ...
Страница 186: ...Figure 5 13 Receiver Gain Adjustment Location 5 24 Adjustments ...
Страница 190: ...Figure 5 16 Receiver Flatness Adjustment Setup 1 MHz 5 28 Adjustments ...
Страница 194: ...Figure 5 20 DC Bias Adjustment Setup 2 5 32 Adjustments ...
Страница 196: ...Figure 6 1 Troubleshooting Organization 6 2 Troubleshooting ...
Страница 206: ...Figure 7 1 Power Supply Lines Simplified Block Diagram 7 2 Power Supply Troubleshooting ...
Страница 212: ...Figure 7 5 A1 CPU Connector Locations 7 8 Power Supply Troubleshooting ...
Страница 220: ...Figure 8 1 Digital Control Group Simplified Block Diagram 8 2 Digital Control Troubleshooting ...
Страница 240: ...Figure 10 1 Top View Major Assemblies 10 4 Replaceable Parts ...
Страница 292: ...Table A 2 Manual Changes by Firmware Version Version Make Manual Changes A 2 Manual Changes ...
Страница 303: ...Change 6 Change the Replaceable Parts as following Figure A 10 Top View Major Assemblies Manual Changes A 13 ...
Страница 308: ......
Страница 311: ...Figure B 1 Power Cable Supplied Power Requirement B 3 ...
Страница 312: ......
Страница 342: ......