Q00180 Rev B
Page 93 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
NETWORK
Sub-Menu
4.
Using the alpha-numeric keyboard, enter in the required TCP/IP network configuration
values/settings in the appropriate fields, as specified in the information previously provided in
the “
Pre-Installation Site Survey Form
”, by the medical facility’s IT System Administrator.
5.
Using the
Menu Control
, arrow down to backlight
Apply
selection in the main menu, and
press the SELECT button
to save the serial port configuration setup, and exit.
8.5.1
“Wireless Network” Equipment - Setup
ZONARE supports wireless connectivity using the
Quatech
(brand)
AirborneDirect
(model)
Wireless Ethernet Bridge (supported model is the ABDG-ET-DP101). This is a plug-and-play wireless
networking solution that is compatible with all ZONARE ultrasound systems that support Ethernet
connectivity. A special power cable, however, is required to supply power to the
Quatech
AirborneDirect
device from a USB port on the ZONARE system, for portable operations.
The
Quatech AirborneDirect
Wireless Ethernet Bridge works with all ZONARE software release
revisions; however the 3.5 and subsequent software releases have specific enhancements to provide
user feedback for signal strength and wireless connectivity.
The
Quatech AirborneDirect
Wireless Ethernet Bridge supports WiFi 802.11b/g wireless standards
and connects to the ZONARE system through a 10 Base-T network interface. This device supports
WEP (64/128 bit) and WPA encryption standards, and LEAP for network authentication (LEAP
required the Quatech device to be configured with a static 128 bit key; this is known as "migration
mode" and is not recommended for longterm use due to the static WEP key
requirement). Before using it with the ZONARE system, the
Quatech AirborneDirect
device must
Summary of Contents for Z.one Ultra SP
Page 1: ......