19.3 Troubleshooting
19.3.1 Transducer
Problems
CAUTIONS
Bent, broken, or missing pins on the transducer connector may cause poor image quality,
including possible mirror image artifact. Be sure to check pins before connecting transducer to
the ZONARE ultrasound system. If pins are bent, broken, or missing, do not use the transducer
and call ZONARE Technical Support.
Figure 48: Transducer Connector Pin Damage – (Example)
Inspection of Transducers
1.
Routinely inspect transducers for the following problems.
•
Signs of wear or damage to the housing, cable, etc.
•
Cracks in case or transducer face
•
Cuts or gouges on any part of the transducer, including transducer face, case, cable, and
connector
•
Buckling or bulging of the lens material on the transducer face
•
Damage to the transducer connector, including bent, broken, deformed, or missing pins
2.
Immediately replace a transducer that exhibits any of the above damage symptoms
19.3.2 Power-On
Problems
SmartCart SP (with docked Scan Engine) fails to power on
1.
Ensure that the AC power is connected to a wall outlet.
2.
Ensure that the circuit breaker (located in the lower-left of the rear panel) is pressed down in
the “
1
” (On) position.
10mmm