Q00180 Rev B
Page 47 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
5
SYSTEM UNCRATING & INSTALLATION PROCEDURES
5.1 Product
Shipment
Shipping containers should be inspected for damages, and the “Tip-n-Tell” indicator for signs of
mishandling during shipment. If any problems are found, immediately notify the shipping carrier. All
installation and set-up of equipment should be done following official ZONARE product installation
procedures.
5.2 Electrical
Requirements
Standard grounded (3-Prong) 110-120 VAC, 15 amp, 60Hz. wall outlets (or proper 220-240 VAC
outlets for systems configured for these power requirements) are required for the
Z.ONEULTRA SP
Diagnostic Ultrasound System.
5.3 Environmental and Space Requirements
The
Z.ONEULTRA SP
Diagnostic Ultrasound System may be located anywhere within the facility, as
long as there is adequate power outlets and remote network connectivity (at facilities utilizing a
DICOM print/storage device hook-up).
5.4 Uncrating
5.4.1 SmartCart
SP:
STEPS:
1.
Prior to opening any packaging, inspect the “Tip-n-Tell” indicator (attached to the outside of
the main shipping box) for signs that the
Z.ONEULTRA SP
has been subjected to shock or tilt
conditions during shipment.
2.
Also inspect the shipping container for any visual signs of rough handling or abuse during
shipment.
3.
Make notes of any discrepancies, and immediately report to the shipping carrier and to
ZONARE’s shipping department representative.
4.
Remove all banding straps surrounding the cardboard main shipping container.
5.
Lift off the top cover portion of the main cardboard shipping container, and set aside.
6.
Remove the small boxes (containing the
Z.ONEULTRA SP
transducers and Scan Engine) from
the inner storage box area (resting atop the SmartCart SP).
7.
Lift the inner storage box upwards, and remove it from the main shipping container.
8.
Grasping the sides, slide the cardboard main shipping box upwards, exposing the
Z.ONEULTRA SP
. Continue lifting until the cardboard box can be completely removed, and
set aside.
Summary of Contents for Z.one Ultra SP