Q00180 Rev B
Page 181 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
Error Code Subsystem
Message for Error Log
User Error Message
Action On
Error
0xe00a0414
HV
HV's Crossed
"Internal System Error has occurred"
RESTART
0xe00a0415
HV
High Voltage
"Internal System Error has occurred"
RESTART
0xe00d0000
Imaging
Pipeline Fault Event Occured.
"Internal System Error has occurred"
RESTART
0xe0100401
Router
Video Memory Test Failed
"Internal System Error has occurred"
SHUTDOWN
0xe0100402
Router
DCM LOCK Fault has occurred);
"Internal System Error has occurred"
RESTART
0xe0110401
AcquisCntrl
Post Power on check indicates faulty
power circuit
"Internal System Error has occurred"
RESTART
0xe0110402
AcquisCntrl
Failed to complete programming
sequence
"Internal System Error has occurred"
RESTART
0xe0110403
AcquisCntrl
Failed to complete programming
sequence
"Internal System Error has occurred"
RESTART
19.5 Scan Engine Fan Operation
The operating speeds of the two (2) fans on the Scan Engine are variable; driven by the dynamic
changes in cooling requirements within the Scan Engine. Four (4) dedicated thermal sensor devices,
strategically located at various locations inside the Scan Engine, provide constant temperature status
information to the Processor (PIC device).
In the case where a temperature sensor error condition occurs, both fans will be set to run at
MAXIMUM speed by the Processor.
19.6 SmartCart SP LCD Display Troubleshooting
19.6.1
Image display quality problems
Symptom
Solution
Images look too bright or saturated
Check the
BRIGHTNESS
,
CONTRAST
and
LUMINANCE
settings, using SmartCart SP
display’s pop-up menu. Verify they are set at (or
close to) the factory default setting (indicated by
an asterisk * in the menu)
Completely blank display
Check cable connection at back of LCD display