Q00180 Rev B
Page 205 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
Installation:
STEPS:
1.
Reconnect the Cables to the replacement Power Supply Module.
2.
Route the printer AC power cable into the chassis, and out the front access panel.
3.
Tuck cables back into chassis and position Power Supply Module for installation.
4.
Reinstall the eight (8) Philips head screws. Ensure that the Power Supply Module is securely
connected.
5.
Tilt system back onto the casters.
System Verification:
STEPS:
1.
Reconnect the main AC power cord from the rear of the SmartCart SP or plug into the wall
source.
2.
Place AC circuit breaker, located at the rear of the SmartCart SP, in the “
1
” (ON) position.
3.
Ensure the scan engine is docked in the SmartCart SP. LED 1 will blink quickly and LED 5
will remain on.
4.
Energize the SmartCart SP.
5.
Verify that the LCD display is functioning correctly by watching the boot and verifying that
normal imaging is displayed after fully booted.
6.
Attach a transducer (if not already connected) and enable PW mode.
7.
Turn PW Gain up to obtain a filled in spectral trace with excess noise (this will generate white
noise for the speakers)
8.
Turn the Volume up (Clockwise) and verify that the white noise is audible from both speakers.
9.
If possible – scanning with actual Doppler flow is ideal, but not required.
10.
Verification is complete – return system to normal operation.
Summary of Contents for Z.one Ultra SP
Page 1: ......