Q00180 Rev B
Page 168 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
NOTE: BootLoad mode results in inability for system to check for scanner docking present,
and does not illuminate any of the five system status LEDs on the rear panel of the SmartCart
SP
-
Turn off the SmartCart SP with another push of the On/Off button
5.
Correcting Power Supply Module out of BootLoad mode:
-
Dock the scanner on the SmartCart SP
-
Manually press the power button directly on the scanner, to turn its power on
-
Press Power On/Off button on scanner Scanner Deck, to turn the SmartCart SP on
-
Go to the System Setup->Diagnostic page, and perform the “Upgrade PIC” operation
SmartCart SP – System Power-Up Testing & PIC “Reset”
1.
Testing On/Off power button operation, and forcing a “PIC Reset” on Power Supply Module:
-
Undock the scanner from the SmartCart SP
-
Press, and hold (for approx. 10 seconds) the Power On/Off button on SmartCart SP
-
LED #1 should blink rapidly and LED #2 should blink at a slower rate, to indicate the PIC is
“Resetting”
1)
IF LED #1 does not blink rapidly, a problem with the Power On/Off button or cabling is
indicated
SmartCart SP – Verifying Basic Power Supply Functionality
−
Verifying for VSCANNER power supply output:
1.
Eject the battery pack from the Scan Engine, and dock scan engine on SmartCart SP
2.
All four of the battery charge LEDs on the Scan Engine should illuminate for approx 5
seconds, and then go out.
3.
LED #5 (VSCANNER) on back of SmartCart SP should be illuminated
---------------------------------------------------------------------------------------------------------------
−
If LED #5 does
NOT
turn on, problem is likely w/Scanner detect circuit or power supply
1.
Press the Power On/Off button on SmartCart SP to check response. If Scan Engine is not
detected, VMAIN (LED #4) won't illuminate. If VMAIN (LED #4) does turn on,
VSCANNER voltage is defective.
---------------------------------------------------------------------------------------------------------------
−
If LED #5
DOES
turn on:
1.
Press Scanner's (not Cart's) Power “On/Off button. Scan Engine (with internal battery still
removed) should boot up, verifying that VSCANNER supply is OK.