Q00180 Rev B
Page 178 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
19.4.2 Scan
Engine
.................................................................................
*
Power-On Switch LED
|
...........................................................
**
Scan Engine Activity LED
: (Solid): Scan Engine is currently powered On
: (Off): Scan Engine is currently powered down
***
Battery Charge Level
..........................................................................
75-100% charged
.............................................................................. 50-75% charged
................................................................................... 25-50% charged
....................................................................................... ****5-25% charged
*POWER-ON SWITCH LED – STATUS INFORMATION:
The Power-On Switch LED is directly activated by the PIC (Controller) device on the Power Board
in the Scan Engine, upon detection of the pressing of the Power On/Off switch. Lack of
illumination of this LED (in conjunction with the Scanner not powering on), indicates one of the
following:
1.
Total loss of source power (8.4V) to the Scan Engine, due to a completely depleted
battery power, or no power from SmartCart SP - if docked.
2.
Power On/Off switch on Scan Engine is defective
3.
“Always On” power output from Power Board in Scan Engine has failed
4.
Low-level
Bootloader
software is corrupted (requires Scan Engine replacement in
field)
Summary of Contents for Z.one Ultra SP