Q00180 Rev B
Page 180 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
19.4.3 System Error Message Codes
Error Code Subsystem
Message for Error Log
User Error Message
Action On
Error
0xa0060402
I2C
ThermalControl Sensor Error
"Thermal Sensor Error: Please shutdown
system." STOP
0xa0060403
I2C
System Overhead Warning
"System Overheat Warning: Finish exam
and shutdown."
CONTINUE
0xa0060405
I2C
Battery Overheat Warning
"Battery Overheat Warning: finish exam
replace battery"
CONTINUE
0xe0004005
n/a
Fatal Error due to NVASSERT
"Internal System Error has occurred"
RESTART
0xe0060401
I2C
Fan is stuck.
"System fan Error: Please shutdown
system." SHUTDOWN
0xe0060404
I2C
Exceeded maximum system error
count.
"System Temperature Limit exceeded."
SHUTDOWN
0xe0060406
I2C
Battery temperature error
"Battery Temperature Limit Exceeded.
Replace battery."
SHUTDOWN
0xe0080001
Display
DVI configuration failed
"Internal System Error has occurred"
SHUTDOWN
0xe0090401
FE Power
Low Voltage failure occurred.
"Internal System Error has occurred"
SHUTDOWN
0xe00a0401
HV
PIC is Invalid, No HV support);
"Internal System Error has occurred"
RESTART
0xe00a0403
HV
Invalid the power up parameters);
"Internal System Error has occurred"
RESTART
0xe00a0404
HV
Failed to reset the PIC HV task);
"Internal System Error has occurred"
RESTART
0xe00a0405
HV
Failed to set HV limits to PIC)
"Internal System Error has occurred"
RESTART
0xe00a0406
HV
Failed to set QDAC voltage to target);
"Internal System Error has occurred"
RESTART
0xe00a0407
HV
Obsolete - See more granual
messages
n/a
n/a
0xe00a0408
HV
Can't Read voltage from PIC);
"Internal System Error has occurred"
RESTART
0xe00a0409
HV
HighVoltage not dumped prior to power
up);
"Internal System Error has occurred"
RESTART
0xe00a040a
HV
VBoost error);
"Internal System Error has occurred"
RESTART
0xe00a040b
HV
PFE HW Fault Error);
"Internal System Error has occurred"
RESTART
0xe00a040c
HV
PIC version is invalid, No HV support);
"Internal System Error has occurred"
RESTART
0xe00a040e
HV
Power Errors (Over Limit)
"Internal System Error has occurred"
RESTART
0xe00a040f
HV
HV Under Voltage
"Internal System Error has occurred"
RESTART
0xe00a0410
HV
HV Over Voltage
"Internal System Error has occurred"
RESTART
0xe00a0411
HV
HV Imbalance
"Internal System Error has occurred"
RESTART
0xe00a0412
HV
HV Power-Up Failed
"Internal System Error has occurred"
RESTART
0xe00a0413
HV
HV Unknown State
"Internal System Error has occurred"
RESTART
Summary of Contents for Z.one Ultra SP