What to do in the event of malfunctions
OPMI Lumera
®
i on floor stand
Version 9.2
Page 130
G-30-1720-en
Zoom error. The error can
be acknowledged.
If the error impairs your
work, please switch to the
manual mode and contact
ZEISS Service.
–
e.g. jammed hardware
If the error impairs your
work, switch to the
manual mode.
Focus error. The error can
be acknowledged.
If the error impairs your
work, please switch to the
manual mode and contact
ZEISS Service.
–
e.g. jammed motor of
the focus system
If the error impairs your
work, switch to the
manual mode.
System boot error. Please
reboot the system. Please
make sure that no function
buttons are pressed during
the booting process.
–
e.g. a button was
pressed during system
start
Reboot the system. Make
sure not to press any
buttons during this
process.
-
System error. The error can
be acknowledged.
If the error recurs, please
contact ZEISS Service.
–
Unknown software or
hardware error.
Acknowledge the error
and contact ZEISS Service
if it occurs again.
-
Error message
Possible cause
Remedy
Reference
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...