OPMI Lumera
®
i on floor stand
Care and maintenance
Version 9.2
G-30-1720-en
Page 133
Safety check
To prevent any impairment of the system’s safety as a result of ageing, wear,
etc., the operating company must ensure, in accordance with the applicable
national regulations, that the regular technical safety checks defined for this
device are performed on schedule and to the stipulated extent.
The safety checks must be performed only by the manufacturer or qualified
personnel.
The scope of the system safety checks should include at least the following:
–
Availability of the instructions for use
–
Visual inspection of the device and its accessories for damage, as well as
legibility of markings and labels
–
Leakage current test
–
Test of protective ground conductor
–
Function and wear test of the steerable casters and locking tabs
–
Functional check of all switches, buttons, connectors, and indicator lamps
on the device
–
Function test of manual mode without the patient, at least semi-annually
CAUTION
Risk of injury without safety inspection!
Without a safety inspection, hazards and device deficiencies will not be
detected in time and can have a negative effect on patients, users, or others.
•
Have a safety inspection performed in compliance with IEC 62353 on
time and to the prescribed extent. Observe the applicable national regu-
lations.
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...