Care and maintenance
OPMI Lumera
®
i on floor stand
Version 9.2
Page 134
G-30-1720-en
Changing the halogen lamp
To change the lamp, proceed as follows:
•
Turn off the device with the power switch.
•
Press button (2) to eject lamp module (1) partially. Pull out the lamp
module.
•
Remove the defective halogen lamp from the spring-loaded mount.
•
Pull the ceramic base (8) from the contact pins of the halogen lamp.
•
Plug the ceramic base (8) onto the contact pins of the new halogen lamp.
•
Insert the new halogen lamp. Make sure you do not touch the lamp
bulb (7) or the interior of the reflector (6).
•
Press the halogen lamp into the spring-loaded mount (4)/(5).
•
Push the lamp module including the new halogen lamp back into the
device.
•
Turn on the device with the power switch.
3
“Instructions for use” sign
Observe the instructions for use or further applicable documents.
9
“CAUTION” warning label
Warning of a potential hazard.
•
Follow the instructions to change the lamp.
NOTICE
10 Risk of burns from hot lamp!
If you replace the lamp shortly after it has failed, the lamp will still be very
hot.
•
Allow the lamp to cool down before removing it.
•
Always use heat-resistant protective gloves when replacing the lamp!
NOTICE
Use the correct halogen lamp!
•
Only use the 12 V, 100 W halogen lamp specified in “Ordering data,
Consumables”.
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...