CAUTION
Never leave the device unattended when it is switched on!
If a patient or an object is exposed to the light source for an extended period
of time, this can cause burns or fire damage. Excessive radiation exposure
times may lead to retinal damage in the patient’s eye.
•
Never leave a device unattended when its light source is switched on.
NOTICE
Restricted functions due to malfunctioning motor electronics!
If the motor electronics malfunction, major functions such as XY movement,
focus, zoom, or light control may be impaired.
•
Activate the manual mode (see page 125).
•
Move the microscope manually into the required position.
•
Set the magnification on the manual zoom adjustment knob.
•
Use an OR illuminator.
•
Inform ZEISS Service.