Operation
OPMI Lumera
®
i on floor stand
Version 9.2
Page 96
G-30-1720-en
•
Check eyepieces / binocular tube / objective lens
The surgical microscope and the tube are in a position convenient
for you (see page 85).
The interocular distance (pupillary distance) has been set
(see page 87).
The eyecups have been adjusted to ensure that you can see the full
field of view (see page 87).
The correct prescription has been set on the diopter scale (see
page 87).
The image quality is correct throughout the entire magnification
range.
The objective lens and eyepieces are clean.
•
Check accessories
The additional equipment (illumination system, video system, etc.)
has been checked for proper function according to the respective
instructions for use.
•
Check correct balancing
When fully balanced, the surgical microscope should remain stationary
in all positions within the work area and neither move upward nor
downward (see page 77).
•
Check downward travel limit
The minimum working distance (height) from the surgical field has been
adjusted (see page 84).
CAUTION
Variable light intensity due to a faulty light source!
A faulty light source with variable light intensity can damage the patient’s
retina.
•
Check the light source and the brightness adjustment before each use
of the device.
NOTICE
Light source failure caused by ageing of the lamp!
As the light source ages, the actual illumination intensity decreases.
The light source intensity may vary or fail.
•
Check each light source before use to make sure it is functioning
properly.
•
Keep a backup lamp or an OR illuminator ready.
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...