Connector panel on the floor stand
1
Strain relief device
for securing power cord and video connection cable.
2
Potential equalization connector
3
Foot control panel connector
4
Remote port
for controlling external devices with a maximum breaking capacity
of 24 V/0.5 A.
5
Indicator window and sliding switch for rated voltage
The rated voltage displayed here must match the rated voltage of the
mains power available at the installation site. You can adjust the sliding
switch using a suitable tool.
6
Power input socket
7
Power output socket
for medical devices of max. 500 VA.
CAUTION
Danger! Electrical voltage!
Only connect accessories and medical devices designated by ZEISS for use
with this system to the power outlet sockets.
•
When connecting other devices, make sure that safety is guaranteed
regarding admissible touch currents and ground leakage currents accord-
ing to IEC 60601-1.
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...