![background image](http://html1.mh-extra.com/html/zeiss/opmi-lumera-i-on-floor-stand/opmi-lumera-i-on-floor-stand_instructions-for-use-manual_3358119128.webp)
What to do in the event of malfunctions
OPMI Lumera
®
i on floor stand
Version 9.2
Page 128
G-30-1720-en
Light guide damaged or
incorrectly connected.
Check that light guide is
correctly seated and replace
if necessary. Contact ZEISS
Service.
Lamp is not properly seated
in the lamp socket.
Properly push lamp into
lamp socket.
Filter selection knob is in an
intermediate position.
Turn knob until the filter is
correctly in place.
Main microscope is not
in the correct working
position.
Reposition the main
microscope.
Main microscope axis
should coincide with optical
axis of patient’s eye.
Image on video monitor
too dark or high level of
noise.
Use of a camera adapter.
Remove fixed stop from
camera adapter.
Instruc-
tions for
use for
camera
Activate the integral light
metering function of the
camera to reduce reflec-
tions from the sclera and
cornea.
Instruc-
tions for
use for
camera
Increase the brightness level
of illumination.
Motor-driven focusing
and/or zoom functions of
surgical microscope are
inoperative.
Failure of stand electronics.
Manually adjust focusing
and/or zoom
on surgical microscope.
Motion of suspension arm
too stiff.
Friction adjustment knob
tightened too firmly.
Slightly loosen friction
adjustment knob.
No video image.
Connection cable not prop-
erly connected.
Check all connections.
-
The stand is not stable.
Floor not level. Stand base
not appropriately posi-
tioned.
Adjust the position of the
stand base.
-
Malfunction
Possible cause
Remedy
Reference
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...