What to do in the event of malfunctions
Version 9.2
G-30-1720-en
Page 125
Failure of a main function (XY, focus, zoom, light)
In the event that one of the major functions fails (XY movement, focus, zoom,
light control), leading to the impairment of further functions, you can switch
to the manual mode to complete a surgical procedure that has already been
started, for example.
•
Press button (1) to switch to the manual mode.
Working in the manual mode
In the manual mode, the light source is set to a medium intensity, and XY
coupling, focus drive, and zoom system are switched off.
•
Manually move the microscope in the X and Y directions for positioning.
•
Manually move the microscope up and down for focusing.
•
Use knob (2) to adjust the zoom value (magnification).
Fig. 36: Switching to the
manual mode
1
2