Preparations for use
OPMI Lumera
®
i on floor stand
Version 9.2
Page 78
G-30-1720-en
1Chip HD Camera option
Installing the 1Chip HD Camera
Please observe the “1Chip HD Camera” instructions for use (G-30-1946).
These instructions for use are included automatically in systems with the
“1Chip HD Camera” option. They contain all the additional explanations for
the “1Chip HD Camera” option.
The 1Chip HD Camera is flanged permanently onto a beam splitter. Normally,
the beam splitter with the 1Chip HD Camera remains installed on the surgical
microscope.
If necessary, the beam splitter with the 1Chip HD Camera can be removed
from the surgical microscope.
CAUTION
Risk of infection due to sagging camera cable!
If the beam splitter and the permanently connected 1Chip HD Camera are re-
moved from the surgical microscope, and the remaining, non-sterile camera
cable is hanging down from the surgical microscope, the camera cable could
contaminate the sterile area with pathogens under certain circumstances.
This may cause infections.
•
Make sure that the remaining camera cable is not hanging down from
the surgical microscope.
R508
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...