Operation
OPMI Lumera
®
i on floor stand
Version 9.2
Page 98
G-30-1720-en
•
Check fundus viewing system (if provided)
When using fundus viewing systems (e.g. BIOM 3 from Oculus), which are
usually installed between the surgical microscope and the patient, make
sure that the patient is neither put at risk nor injured by the motorized
focusing system or the movement of the stand arm.
Only use accessories expressly certified by the manufacturer for combina-
tion with the surgical microscope described in these instructions for use.
CAUTION
Injury to the patient’s eye due to incorrect operation of the fundus
viewing system!
Incorrect operation of the fundus viewing system may cause injury to the
patient and impair the image quality. Perform the following steps to avoid
this hazard:
•
Perform a focus reset.
•
With the fundus viewing system swung out of the beam path, always
position the microscope body in such a way that the index dot (1)
of the microscope’s focus is in the middle of the upper triangle (2)
of the marking.
•
Select a medium magnification (e.g. 1.0).
•
Lower the surgical microscope to the surgical field until the patient’s
cornea is in focus.
•
Turn the locking lever for limiting downward travel clockwise; without
any patients, check that the suspension arm cannot be lowered any
further.
•
It is vital that you read the instructions for use of the fundus viewing
system (e.g. BIOM 3 from Oculus).
1
2
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...