What to do in the event of malfunctions
OPMI Lumera
®
i on floor stand
Version 9.2
Page 126
G-30-1720-en
Troubleshooting
For your safety
•
Many malfunctions are detected automatically. Therefore, pay attention
to the error messages displayed on the 5.7
″
control panel (see page 129).
•
If there is a failure you cannot repair using the instructions from the chap-
ter “What to do in the event of malfunctions”, attach a sign “Out of order”
to the system and inform ZEISS Service.
System malfunctions
Malfunction
Possible cause
Remedy
Reference
No function at all.
Power plug of stand not
inserted.
Plug in the power cord.
Power switch of stand not
switched on.
Press power switch.
Green indicator light in the
power switch must be lit.
Automatic circuit breaker
in the stand power switch
is being triggered.
Press the power switch
again.
Power failure.
Contact in-house
electrician.
-
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...