OPMI Lumera
®
i on floor stand
What to do in the event of malfunctions
Version 9.2
G-30-1720-en
Page 129
Malfunctions of the FCP foot control panel
Error messages on the 5.7
″
control panel
In the event of malfunctions occurring during operation, an error message
is displayed on the 5.7
″
control panel.
•
Check and eliminate the malfunction.
•
Acknowledge the message.
The following table contains examples of error messages and the corre-
sponding remedies. If you have problems eliminating the malfunction,
or if errors keep recurring, do not continue using the system and contact
ZEISS Service.
You will find the contact responsible for your country on the following
website: www.meditec.zeiss.com
Malfunction
Possible cause
Remedy
Reference
Foot control panel is
inoperative.
Connections on stand or
foot control panel not
plugged in.
Plug in connections.
Error message
Possible cause
Remedy
Reference
XY coupling error. The error
can be acknowledged.
If the error impairs your
work, please switch to the
manual mode and contact
ZEISS Service.
–
e.g. hardware error in
motor electronics
–
e.g. software error,
motor control in
forbidden status
If the error impairs your
work, switch to the
manual mode.
Light source error. The error
can be acknowledged.
If the error impairs your
work, please switch to the
manual mode and contact
ZEISS Service.
–
e.g. malfunction of light
intensity setting
If the error impairs your
work, switch to the
manual mode.
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...