OPMI Lumera
®
i on floor stand
Overview of devices
Version 9.2
G-30-1720-en
Page 53
Control elements on the lamp housing
The lamp housing is designed for fiber illumination and is equipped with a
halogen light source. The lamp housing contains a backup lamp which auto-
matically swings into the illumination beam path when the first lamp fails.
1
Lamp module
2
Flap
for indicating the operating status of the halogen lamps.
–
When the flap is closed, the main lamp is operative.
–
When the flap is open (yellow mark on the side), the
backup lamp is in use.
3
Manual switching to the backup lamp
4
Button for opening the lamp module
5
Button for setting the filter
Fig. 13: Control elements on
the lamp housing
No filter
Retina protection filter (blue barrier filter): when operating on
the eye, always use this filter to protect the patient’s eye against
unnecessary (blue) radiation.
KK 40 filter:
to increase the color temperature
Empty filter position
1
2
3
4
5
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...