G-30-1720-en
Electromagnetic immunity for ME equipment or systems that are not
life-sustaining
OPMI Lumera i is intended for operation in the electromagnetic environment
specified below. The customer or the user of OPMI Lumera i is responsible for
ensuring that the device is operated in such an environment.
Immunity tests
IEC 60601 test level
Compliance level
Conducted HF disturbances as
per IEC 61000-4-6
3 V
150 kHz to 80 MHz
3V
6 V
ISM frequency bands from
150 kHz to 80 MHz
SIP/SOPs tested with lengths
<3 m
6V
Radiated HF disturbances as
per IEC 61000-4-3
3 V/m
80 - 2700 MHz
3 V/m
Radiated HF disturbances from
near fields of wireless commu-
nication devices as per
IEC 61000-4-3
27 V/m
380 - 390 MHz
27 V/m
28 V/m
430 - 470 MHz,
800 - 960 MHz,
1700 - 1990 MHz,
2400 - 2570 MHz
28 V/m
9 V/m
704 - 787 MHz,
5100 - 5800 MHz
9 V/m
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...