000000Ć1150Ć839 IOLMaster 11.02.2004
Measurement information
Comparison results of IOLMaster versus conventional measurements at the human eye*:
Axial length**
Corneal radius***
Anterior chamber depth**
Mean values of Differences
Standard Deviations
-0.01 mm
$
0.19 mm
-0.01 mm
$
0.05 mm
0.07 mm
$
0.18 mm
IOLMaster measuring reproducibility*:
Axial length
Corneal radius
Anterior chamber depth
Regarding the human eye
regarding the artificial model eye
standard deviation****
standard deviation
$
0.03 mm
$
0.01 mm
$
0.02 mm
$
0.02 mm
$
0.10 mm
$
0.10 mm
*
Data on File
**
Compared to high precision ultrasound immersion device
***
Compared to a handheld keratometer
****
Standard deviation determined on basic of xi" calculated single standard deviations
Technical Data